Main components of the automatic mycotoxin analytical chromatography system Manufacturer: R-Biopharm
1. UHPLC ultra-high-pressure analysis system
- Power supply: 95/230 VAC, 50/60 Hz, max. 150 VA
- Operating noise level: LAeq 70dB
- Can be synchronised with UV, FLD and MS/MS detectors from different manufacturers.
a. Ultra-high-pressure pump of the automatic mycotoxin chromatographic analysis system (RIDA®CREST Gradient pump set):
- Flow rate: 0.001-5.00 ml/min
- Flow rate increment: 0.001 ml.min
- Has a function to change the mobile-phase composition over the analysis time (gradient high pressure).
- Maximum pressure: 124 Mpa (~18000psi)
- Flow-rate repeatability: < 0.2 % RSD (for methanol at 1ml/min, pressure 10000 psi)
- Flow-rate accuracy: < 1 % (for methanol at 1ml/min, pressure 10000 psi).
- Operating pH range: 1 - 10
- Output pressure pulsation: ± 0.75 % (in the range 500psi to 18000 psi)
- Large-volume mixer: 60μL.
- Automatic continuous pump rinsing.
- Integrated solvent degasser (Degasser)
- Working environment: 5-40°C, 20-80 % RH
- Parts in contact with the solvent and the mobile phase are made from: 316 steel, ruby, sapphire, Teflon, UHMWPE, Teflon AF, PEEK.
- Dimensions: 300 x 120 x 500 mm
- Mass: 16 kg
b. Automatic sample injector of the automatic mycotoxin analytical chromatography system (Autosampler):
- Injection mode: Full-loop, Partial loop-fill
- Sampling needle of silicon-coated steel and PEEK TM valve
- Injection volume: 0-5000 µl (depending on the sample loop and injection mode )
- Sample-aspiration syringe volume: 500 µl (standard); 250, 1000, 2500 µl are additional options.
- Sample injector resolution: 1 µl
- Injection-volume accuracy: Full-loop <0.3% RSD, Partial loop-fill <0.5% RSD (At an aspiration volume greater than 5 µl).
- Cross-contamination: 0.01% with the extended wash mode and 0.05% with the standard wash mode.
- Sample viscosity range: 0.1 - 5 cP.
- Number of sample vials: 2 x 48 (1.5 ml), 2 x 12 (10 ml) or 2 plate 96 - wells.
- Needle wash mode: Internal and external needle wash with needle drying.
- Time per wash cycle: < 20s with an injection volume of 10 µl and the loop washed with wash buffer. (time < 60 s in all injection modes with injection volume ≤ 100 µl and washing with 300 µl)
- Injection valve switching time: 60 msec
- Cooling unit: the maximum cooling temperature is 4°C and the maximum cooling temperature is 3°C below the ambient temperature.
- Parts in contact with the mobile phase and the solvent are made from: SS316, PTFE, TeFzel, Vespel, glass, Teflon. For the Bio kit: PEEK and coated steel including SS316.
- Mass: 21kg
- Dimensions: 300 x 575 x360 mm
- Operating temperature: 10°C - 40°C, humidity 20% - 80% RH
c. Column oven of the automatic mycotoxin analytical chromatography system (RIDA®CREST column oven):
- Capacity: 6 analytical columns ( column length 250mm)
- Standard automatic column-switching unit for HPLC ( UHPLC is an additional option)
- Controllable temperature range: 5 - 75°C
- Temperature accuracy: ± 0.1°C (measured at 30°C)
- Temperature repeatability: ± 0.1°C (measured at 30°C)
- Temperature change time: 10°C/min from 40°C up to 60°C
- Cooling time: 2°C/min from 60°C to 40°C
- Optional temperature change time: up to 9h59p with a 1-minute step.
- Temperature acceleration parameter when increasing temperature: 0.1 – 5.0°C.
- Temperature acceleration parameter when decreasing temperature: 0.1 – 1.5°C/min (from 75°C down to 25°C)
- Dimensions: 600 x 170 x 345 mm
- Weight: 16 kg
2. RIDA®CREST High Pressure Dispenser high-pressure liquid dispensing system (supplied with the automatic sample clean-up unit with the clean-up-column self-checking function):
- High-pressure syringe, volume 2ml. Made of stainless steel
- Flow rate : 0.1 - 10 ml/min
- Flow rate accuracy: < 2 %
- Volume range: 0.1 ÷ 1 ml, resolution 0.01 ml
- 0.1-1 ÷ 10 ml, resolution 0.1 ml
- 10 ÷ 100 ml, resolution 1 ml
- Volume error: < 1 %
- Pressure range: 0 - 300 bar. Independent of the ultra-high-pressure pump.
- Number of solvent selection boxes: 2 SSMs for 13 solvents.
- pH range: 1.0 - 12.5
- Function: solvent dispensing for the automatic sample cleanup unit.
- Sequential operation according to the control software.
- Interface: 2 TTL digital inputs and 2 outputs 28V-1A
- Software: Clarity 5.04
- Power supply: 115/230 VAC, 50/60 Hz, max.150 VA
- Dimensions: 150 x 285 x 355 mm
- Mass: 11 kg
- Working environment: 5 - 40°C, 20 - 80 % RH
3. Self-checking sample clean-up unit using RIDA®CREST immunoaffinity micro-columns Automated Cartridge Exchange:
- Cartridge capacity: 2 cartridge trays, each tray 96 immunoaffinity columns.
- Affinity cleanup column: microcolumn, length 1 cm.
- Number of reuses for 1 cleanup column: 15 times.
- The control information system includes: identification number for the immunoaffinity column, immunoaffinity column type used, batch number, production number and expiry date.
- Cartridge tray: 8 x 12 wells (96 x 128 mm)
- Cartridge lock: no leakage up to 300 bar
- High-pressure valve: 6-port valve
- Cartridge replacement cycle: < 13 s
- Interface: 4 TTL digital inputs and 4 outputs 28V-1A
- PC communication: RS232C
- Software: Clarity 5.04 (Eser)
- Power supply: 115/230 VAC
- Dimensions: 300 x 640 x 360 mm
- Weight: 19 kg
- Working environment: 5 - 40°C, 20 - 80 % RH
4. Probe (Supplied by the customer depending on the intended use)